Selasa, 22 Maret 2016

HANNA HI84531 Titratable Alkalinity Mini Titrator for Water Analysis


The HI84531 is an easy to use, fast and affordable automatic mini titrator designed for testing titratable alkalinity levels in water. Based on an acid base titration method, this mini titrator uses an optimized pre-programmed method of analysis with a powerful algorithm that determines the completion of the titration reaction by the use of a glass body pH electrode. The HI84531 incorporates a precision piston style dosing pump that adjusts the volume of dosing dynamically based on the voltage change. This dosing system reduces the amount of time required for the titration while providing for a highly accurate determination of the amount of titrant used. The HI84531 is supplied complete with all the materials necessary to perform low and high range measurements of alkalinity in water. All chemicals are premixed and prepackaged including standardized titrants, reagents and pump calibration solution. There is no need for volumetric glassware or analytical balances.

Piston driven pump for accurate dynamic dosing

HI1131B double junction pH electrode

Pre-standardized titrant and pre-measured reagents

Details

Total titratable alkalinity is a measure of primarily three types of alkalinities present in a water sample: hydroxide, carbonate and bicarbonate. Alkalinity in water can be the result of contributions from common chemicals including carbonates, bicarbonates, hydroxides, phosphates, borate and organic acid salts.

The alkalinity of a water sample indicates its ability to resist pH change. The amount of alkalinity in water is mostly due to the bicarbonate/carbonate present. A low alkalinity level indicates that the water is susceptible to pH changes, while a high alkalinity level indicates that the water will be able to resist pH changes. Alkalinity can also be used to determine the corrosive capacity of water and can provide an estimation of water hardness.

Double Junction pH Electrode

The HI84531 is supplied with the HI1131B refillable, double junction, combination pH electrode. By design, the HI1131B has a spherical tip for use in aqueous or liquid solutions. This versatile electrode provides a wide surface of contact with a sample and is ideal for direct measurements or acid base titrations in the water industry.

Piston Driven Dosing Pump

The heart of the HI84531 is the piston driven burette pump. This type of dosing system uses a motor in which each dose is very precisely controlled and the volume dispensed is accurately determined. The piston driven burette is controlled dynamically so that the volume of titrant being dosed is automatically adjusted based on a voltage response of the previous dose. This type of dosing speeds up the titration process by allowing for more titrant to be dosed at the start of the titration and then very small doses as the endpoint is reached.

Automatic Stirrer

The built-in stirrer is automatically maintained at a speed of 600 rpm, regardless of the viscosity of the solution being titrated.

On-screen Features

titrator screenshot
Easy-to-use Interface The HI84531 has an intuitive user design with clearly defined keys and a large screen that is easy to navigate. The meter has a built in tutorial mode that, when enabled, will walk the user step by step through the titration process. A dedicated HELP key is always available to allow access to content-specific information during both calibration and titration.
Titrate LR
Procedure Warnings Users are warned if there is an error in procedures such as the titration exceeded the maximum volume of titrant.
mini titrator acidity graphic mode
Graphic Mode This mini titrator displays in-depth data during titration, including a real-time graph of the titration curve.
mini titrator acidity logging
Log-on-demand The HI84531 allows for data logging of up to 400 samples: 200 titration results and 200  pH/mV readings. Data can be stored and exported to a USB drive or a PC using the USB connection.
mini titrator pH mv meter
pH/mV Meter In addition to being an automatic titrator, the HI84531 also can be used as a pH/mV meter. As a pH meter, it has many features of a professional grade benchtop including automatic calibration up to three points with four available buffers, a 0.01 pH resolution, accuracy of ±0.01 pH, automatic temperature compensation and comprehensive GLP data.  
mini titrator acidity graphic mode
CAL Check Accuracy is always ensured with Hanna’s unique CAL Check™ feature, which analyzes the response of the electrode during the calibration process. Based on electrode response in the buffer, indicators are displayed on screen to alert the user of potential problems during calibration. These indicators include Buffer Contaminated, Electrode Dirty/Broken, and overall probe condition. The CAL Check™ function not only ensures an accurate pH reading when the HI84531 is used as a pH meter, but also an accurate titration since the endpoint of a titratable acidity titration is determined by a set pH value.

Specifications

RangeLow Range: 30.0 to 400.0 mg/L; 0.6 to 8.0 meq/L; High Range: 300 to 4000 mg/L; 6.0 to 80.0 meq/L
ResolutionLow Range: 0.1 mg/L (ppm); 0.1 meq/L; High Range: 1 mg/L (ppm); 0.1 meq/L
Accuracy (@25ºC/77ºF)Low Range: ±1 mg/L or 3% of reading, whichever is greater; High Range: ±10 mg/L or 3% of reading, whichever is greater
Sample VolumeLow Range: 10 mL; High Range: 2 mL
Methodsacid-base titration (strong alkalinity/total alkalinity)
Principleendpoint titration: 8.30 pH (phenolphthalein) / 4.50 pH (bromcresol green-methyl red)
Pump Speed10 mL/min
Stirring Speed600 rpm
pH Range-2.0 to 16.0 pH; -2.00 to 16.00 pH
pH Resolution0.1 pH / 0.01 pH
pH Accuracy (@25ºC/77ºF)±0.01 pH
pH Calibrationone, two or three-point calibration; four available buffers (4.01, 7.01, 8.30, 10.01)
Temperature Compensationmanual or automatic
mV Range-2000.0 to 2000.0 mV
mV Resolution0.1 mV
mV Accuracy (@25ºC/77ºF)±1.0 mV
Temperature Range-20.0 to 120.0°C; -4.0 to 248.0°F; 253.2 to 393.2 K
Temperature Resolution0.1°C; 0.1°F; 0.1K
Temperature Accuracy (@25ºC/77ºF)±0.4°C; ±0.8°F; ±0.4 K
Logging Dataup to 400 ( 200 titrations, 200 pH/mV readings)
ElectrodeHI1131B glass body, refillable, with BNC connector and 1 m (3.3’) cable (included)
Temperature ProbeHI7662-T stainless steel temperature probe with 1 m (3.3’) cable (included)
Environment0 to 50°C (32 to 122°F); RH max 95% non-condensing
Power Supply12 VDC adapter (included)
Dimensions235 x 200 x 150 mm (9.2 x 7.9 x 5.9”)
Weight1.9 kg (67.0 oz.)
Ordering InformationHI84531-01 (115V) and HI84531-02 (230V) are supplied with HI1131B pH electrode, HI7662-M temperature probe, HI84531-50 low range titrant (120 mL), HI84531-51 high range titrant (120 mL), HI84530-55 calibration standard (230 mL), 100 mL beakers (2), dosing pump valve, 5 mL syringe, 1 mL plastic pipette, tube set (aspiration tube with titrant bottle cap and dispensing tube with tip), stir bar, power adapter, instruction manual and quality certificate.